Features
High-speed sampling and wide range measurement The power of devices driven at a high frequency can be measured at a 100 MS/s sampling rate, at a 12-bit resolution, and in the 20 MHz range*1 . |
Waveform measurement function Instantaneous power waveforms can be displayed as standard in addition to voltage and current waveforms, and power changes can be observed directly. Voltage, current, and power waveforms for each cycle can be calculated and numerical values can be displayed by cursor. The average voltage, current, and power values in a specified period by the cursor can be calculated. Acquisition memory is up to 100 M points per channel (when equipping the /M2 option), allowing for capturing and displaying detailed waveforms. |
Waveform analysis function Up to the 500th order harmonic components can be measured simultaneously (when installing the /G5 option). 2-channel FFT function is available as standard |
De-skew (phase compensation) function when using an external current sensor, etc. is available. |
Motor characteristics can be evaluated (mechanical output calculation with torque and rotation speed input, as well as analog and pulse input). |
Specifications
Current direct input range: 10 m/20 m/50 m/100 m/200 m/500 m/1/2/5 Arms |
Current sensor input range: 50 m/100 m/200 m/500 m/1/2/5/10 Vrms |
Frequency range: DC to 20 MHz (-3 dB, voltage and current sensor input) DC to 10 MHz (-3 dB, current direct input) |
Power basic accuracy (45 Hz to 66 Hz): ±(0.1% of reading + 0.1% of range) |
Influence of power factor error (cos ϕ = 0): ±0.15% of S (apparent power) |
A/D converter: Maximum sampling rate 100 MS/s Resolution 12-bit |
Acquisition memory: Standard: 10 M points per channel Max: 100 M points per channel (/M2 option) |
Maximum waveform viewing time: 20 minutes (not dependent on the memory size) |
History memory: This function can save up to 1000 records of waveform data and display and calculate them as needed (when the /M2 option installed) |
Waveform display: Up to 16 waveforms can be displayed. Voltage and current waveforms and simultaneous power waveform can be displayed. |
Snapshot: Waveform at an arbitrary moment on the screen can be saved. |
De-skew (phase compensation) function: Phase difference between the voltage and current modules is compensated. |
Trend measurement (waveform measurement, MATH): Voltage, current, and power waveform calculation for each cycle |
Calculation in the specified period (waveform parameter calculation, MEASURE): Average value between cursors can be measured |
Simultaneous harmonic measurement: Up to the 500th order harmonic measurement (/G5 option) |
2-channel FFT function available as standard |
Printer: Screens can be copied (/B5 option) |
External storage: USB port (×2), SD card |
Video output: RGB analog, video output |
Display unit: 10.4-inch color TFT XGA display |
GP-IB, Ethernet, and USB communication available as standard |
IRIG function: Data measured with multiple PX8000 units can be synchronized (/C20 option) |
Sensor power supply: 4CH DC power supply ±15Vdc Max. of 1.8A/CH |
External dimensions: 355 (W) × 259 (H) × 180 (D) mm (excluding protrusions) |
Weight: Approximately 6.5 kg (main unit only, excluding paper and options) |