GENERAL
|
|
Brand | IET Labs, Inc |
---|---|
Sub Categories | Lab tools |
Application | Standard Capacitor |
|
|
The SCA Standard Capacitor series are highly stable cost-effective capacitance standards with low temperature coefficient, low losses and a wide range of values of 1 pF – 10,000 µF.
The cost and flexibility make this our most popular capacitance standard.
New SCA-1nF-8588A
IET Labs has designed the SCA-1nF-8588A to calibrate the Fluke 8588A 1 nF capacitance range as discussed in the Verifying the capacitance function of the 8588A Reference Multimeter Applications Page.
The 1 nF range in the Fluke 8588A is the only range that does not use a resistance range and can be checked/calibrated with the SCA-1nF-8588A. The other capacitance ranges can be monitored by observing the change in the appropriate resistance range.
(Full Specifications can be found on the datasheet)
Calibration:An accredited certificate of calibration is provided with each unit
Nominal value
|
Model
|
Adjustment to nominal
|
Temperature coefficient (ppm/°C)
|
Calibration* frequency
|
Dissipation (typical)
|
Stability(per year)
|
Terminals
|
Capacitor Type
|
1 pF
|
SCA-1pF
|
±0.1%
|
+20 to +40
|
1 kHz
|
0.002
|
±0.1 pF
|
2 bnc + gnd
|
Air capacitors
|
1.9 pF
|
SCA-1.9pF
|
±0.1%
|
+20 to +40
|
1 kHz
|
0.002
|
±0.1 pF
|
2 bnc + gnd
|
Air capacitors
|
10 pF
|
SCA-10pF
|
±0.1%
|
+20 to +40
|
1 kHz
|
0.002
|
±0.1 pF
|
2 bnc + gnd
|
Air capacitors
|
19 pF
|
SCA-19pF
|
±0.1%
|
+20 to +40
|
1 kHz
|
0.002
|
±0.1 pF
|
2 bnc + gnd
|
Silvered Mica
|
100 pF
|
SCA-100pF
|
±0.1 pF
|
20
|
1 kHz
|
0.0005
|
±100 ppm
|
2 bnc + gnd
|
Silvered Mica
|
190 pF
|
SCA-190pF
|
±0.1 pF
|
20
|
1 kHz
|
0.0005
|
±100 ppm
|
2 bnc + gnd
|
Silvered Mica
|
1 nF
|
SCA-1nF
|
±0.02%
|
20
|
1 kHz
|
0.0003
|
±100 ppm
|
2 bnc + gnd
|
Silvered Mica
|
1.9 nF
|
SCA-1.9nF
|
±0.02%
|
20
|
1 kHz
|
0.0003
|
±100 ppm
|
2 bnc + gnd
|
Silvered Mica
|
10 nF
|
SCA-10nF
|
±0.02%
|
20
|
1 kHz
|
0.0003
|
±100 ppm
|
2 bp’s + gnd
|
Silvered Mica
|
19 nF
|
SCA-19nF
|
±0.02%
|
20
|
1 kHz
|
0.0003
|
±100 ppm
|
2 bp’s + gnd
|
Silvered Mica
|
100 nF
|
SCA-100nF
|
±0.02%
|
20
|
1 kHz
|
0.0003
|
±100 ppm
|
2 bp’s + gnd
|
Silvered Mica
|
190 nF
|
SCA-190nF
|
±0.02%
|
20
|
1 kHz
|
0.0003
|
±100 ppm
|
2 bp’s + gnd
|
Silvered Mica
|
1 μF
|
SCA-1μF
|
±0.02%
|
20
|
1 kHz
|
0.0002
|
±100 ppm
|
2 bp’s + gnd
|
Silvered Mica
|
1.9 μF
|
SCA-1.9μF
|
±0.02%
|
20
|
1 kHz
|
0.0002
|
±100 ppm
|
4 bp’s + gnd
|
Silvered Mica
|
5 μF
|
SCA-5μF
|
±0.02%
|
±50
|
1 kHz
|
0.0005
|
±200 ppm
|
4 bp’s + gnd
|
MPPS
|
10 μF
|
SCA-10μF
|
±0.04%
|
±50
|
100 Hz
|
0.0005
|
±200 ppm
|
4 bp’s + gnd
|
MPPS
|
19 μF
|
SCA-19μF
|
±0.04%
|
±50
|
100 Hz
|
0.0005
|
±200 ppm
|
4 bp’s + gnd
|
MPPS
|
100 μF
|
SCA-100μF
|
±0.05%
|
±50
|
100 Hz
|
0.001
|
±500 ppm
|
4 bp’s + gnd
|
MPPS
|
190 μF
|
SCA-190μF
|
±0.05%
|
±50
|
100 Hz
|
0.001
|
±500 ppm
|
4 bp’s + gnd
|
MPPS
|
1,000 μF
|
SCA-1000μF
|
±0.4%
|
-150
|
100 Hz
|
0.001
|
±500 ppm
|
4 bp’s + gnd
|
MPPS
|
5,000 μF
|
SCA-5000μF
|
±2%
|
-150
|
100 Hz
|
0.001
|
–
|
4 bp’s + gnd
|
Polypropylene
|
10,000 μF
|
SCA-10000μF
|
±2%
|
-150
|
100 Hz
|
0.001
|
–
|
4 bp’s + gnd
|
Polypropylene
|