Features
Excellent basic performance polished to details Basic power accuracy: ± (0.01% of reading + 0.02% of range) Measurement bandwidth: Voltage DC to 10 MHz, Current DC to 5 MHz Capture a slight value change in various condition of motor drive |
Functions to support simultaneous, synchronized measurement of multi-input systems Simultaneous power measurement of up to 7 inputs Evaluation of up to 4 motors (optional) WXGA LCD with touch screen USB (3.0)/Ethernet/GP-IB communications (standard) Up to 32 GB of non-volatile internal memory (optional) |
Improved measurement performance for inverters Max. 10 MS/s & 18 bits AD converter equipped Simultaneous dual harmonic measurement up to the 500th order (standard) Harmonic analysis of fundamental waveform up to 300 kHz High-accuracy simultaneous measurement of power parameters of an entire bandwidth and fundamental components Various Line/Frequency filter functions for detail analysis |
Specifications
Voltage ranges: 1.5/3/6/10/15/30/60/100/150/300/600/1000 V |
Current ranges (Direct input): 0.5/1/2/5/10/20/30 A (760901) 5 m/10 m/20 m/50 m/100 m/200 m/500 m/1/2/5 A (760902) |
Current ranges (External current sensor input): 50 m/100 m/200 m/500 m/1/2/5/10 V |
Measurement bandwidth (Power): DC, 0.1 Hz to 1 MHz |
Basic power accuracy (45 Hz to 66 Hz): ± (0.01% of reading + 0.02% of range) |
DC power accuracy: ± (0.02% of reading + 0.05% of range) |
Date update rate: 50 m/100 m/200 m/500 m/1/2/5/10/20 s |
Effect of Power factor: ±0.02% of S (S: Apparent power at cos ϕ =0 ) |
A/D converter Sample rate: Up to 10 MS/s Resolution: 18 bits |
Display: 10.1 inch Color TFT (WXGA) Touch screen |
Communication I/F: Standard function GP-IB, Ethernet (1000Base-T, VXI-11)and USB (3.0 USB-TMC) |
External dimensions: Approx. 426(W) × 177(H) × 469(D) mm |
Weight: Approx. 12.5 kg (Main frame without input element) |