Ø170mm
• Can measure AC current up to 3000A
• Flexible and light weight by the air core coil used at the sensor part
• Designed to meet the international safety standard IEC 61010-1 CAT IV 300V, CAT III 600V
Applicable model | 5050(*), 6305, 6315 |
---|
* Please check the “Applicable model list” before using the product.
Conductor size | max.Ø170mm |
---|---|
Rated current | AC 3000A |
Output voltage | AC 500mV / 3000A (AC 0.167mV/A) |
Accuracy | ±1.0%rdg±0.5mV (45Hz – 65Hz) ±1.5%rdg±0.5mV (40Hz – 1kHz) |
Phase shift | within ±2.0° (45Hz – 65Hz) within ±3.0° (40Hz – 1kHz) |
Cable length | Approx. 3m |
Output connector | MINI DIN 6PIN |
Output impedance | 100Ω or less |
Applicable standards | IEC 61010-1, IEC 61010-2-032 CAT IV 300V / CAT III 600V Pollution degree 2, IEC 61326 |
Operating temperature & humidity range | -10 – 50°C, relative humidity 85% or less (no condensation) |
Weight | Approx. 200g |
Included Accessories | Instruction manual, Cable marker, 9095 (Carrying case) |