Charge Plate Monitors – Monroe 288C

Thương hiệu : Trek

Cam kết và Chính sách của TECHNO

  • icon

    Cam kết 100% Hàng chính hãng

  • icon

    Chất lượng cao, đáp ứng đa dạng nhu cầu

  • icon

    Tư vấn kỹ thuật toàn diện và chuyên sâu

  • icon

    Chính sách bảo hành và hỗ trợ nhanh chóng

Easy-to-Use Charge Plate Monitor For Manual or Automated Testing

Advanced Energy’s Monroe 288C is the first charged plate monitor to offer onboard data storage to automate ionization system testing. The Monroe 288C performs customary manual tests or a series of automated decay and balance tests.

  • Electrometer Dynamic Range: ±1200 V
  • Charge Plate Size: 15 x 15 cm (6 x 6 in)
  • Data Storage: 1500 Readings

Overview

The MonroeTM 288C is an easy-to-use, charge plate monitor for manual or automated testing. All test parameters are programmable, allowing tests to be optimized and not dictated by equipment limitations.

Once programmed, the Monroe 288C will perform a series of tests automatically: ± decays, balance, balance peaks, temperature, humidity, time/date are stored and may be reviewed via the display or downloaded to a PC. The included software permits the user to define and name ionizer locations, test setups, and sequences and upload them to the CPM. All of these features result in a flexible, easy-to-use instrument that facilitates audits while minimizing errors.

Benefits

  • Perform manual and automated testing of decay and balance
  • Store up to 1500 tests total (not for each location), 600 locations, and 4 test protocols
  • Use the large, easy-to-read, high contrast LCD display
  • Configure operating parameters
  • Utilize soft keys for highly intuitive programming

 

Features

  • Internal battery for portable operation
  • Detachable 6 x 6-inch plate with optional 1 x 1 in and 3 x 3 in plates
  • Built-in temperature and humidity sensors
  • Auto-ranging to 0.1V resolution below 100V Meets ESD Association Standard ANSI / ESD STM 3.1
  • CE compliant
  • Performance

    Range

     

    ±1200 V

     

    Bandwidth

     

    -3db @ 1Khz 20 V p-p
    -3db @ 10Hz 2000 V p-p

     

    Noise

     

    < 12 mV rms

  • Charge Plate

    Capacitance

     

    20 pF ± 2 pF

     

    Resolution

     

    Settable to 1 V increments

     

    Accuracy

     

    0.3% of setting ±2.5 V

  • Mechanical

    Battery Life

     

    Typ > 6 hours

     

    Size

     

    280 x 229 x 152 mm (11 x 9 x 6 in)

     

    Weight

     

    12.5 lb (5.7 kg)