DESCRIPTION
To those who wish to double production efficiency
Just 5 mg after 30 million times!
AD-4212C-300 (320 g × 1 mg) completed an endurance test of 30 million repeated loadings, which took more than 1 year.
The final value of span drift was a mere 5 digits, or 5 mg (display value of 200.005 g for a standard weight of 200 g), demonstrating that the product can maintain its accuracy over a long period of use.
See how the test was performed here
Features | |
Fast stabilization (0.5 sec. at 1 mg and 1.3 sec. at 0.1 mg) Slim size (59 mm width) for installation in narrow spaces Direct digital data output to a panel computer, PLC, or PC via RS-232C High-speed data transmission of up to 50 times/second IP65 dust and waterproof construction Protection mechanism against vertical and transverse overloading due to actuator malfunctions, etc. Removable leveling feet for direct attachment on the mounting base Free WinCT-AD4212C software for easy calibration and response setting Comes with a free calibration weight (OIML Class E2 equivalent) CC-Link connection to a PLC through the optional AD-8923-CC remote controller (AD-8923-BCD for BCD output is also available) |
|
Videos | |
External Input/Output | |
RS-232C interface |